Class PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

java.lang.Object
com.irurueta.navigation.inertial.calibration.accelerometer.RobustKnownBiasAndPositionAccelerometerCalibrator
com.irurueta.navigation.inertial.calibration.accelerometer.PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator
All Implemented Interfaces:
AccelerometerCalibrator, AccelerometerNonLinearCalibrator, KnownBiasAccelerometerCalibrator, OrderedStandardDeviationBodyKinematicsAccelerometerCalibrator, QualityScoredAccelerometerCalibrator

public class PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator extends RobustKnownBiasAndPositionAccelerometerCalibrator
Robustly estimates accelerometer cross couplings and scaling factors using a PROMedS algorithm to discard outliers.

To use this calibrator at least 10 measurements taken at a single known position must be taken at 10 different unknown orientations and zero velocity when common z-axis is assumed, otherwise at least 13 measurements are required.

Measured specific force is assumed to follow the model shown below:

     fmeas = ba + (I + Ma) * ftrue + w
 
Where: - fmeas is the measured specific force. This is a 3x1 vector. - ba is accelerometer bias. Ideally, on a perfect accelerometer, this should be a 3x1 zero vector. - I is the 3x3 identity matrix. - Ma is the 3x3 matrix containing cross-couplings and scaling factors. Ideally, on a perfect accelerometer, this should be a 3x3 zero matrix. - ftrue is ground-truth specific force. - w is measurement noise.
  • Field Details

    • DEFAULT_STOP_THRESHOLD

      public static final double DEFAULT_STOP_THRESHOLD
      Default value to be used for stop threshold. Stop threshold can be used to avoid keeping the algorithm unnecessarily iterating in case that best estimated threshold using median of residuals is not small enough. Once a solution is found that generates a threshold below this value, the algorithm will stop. The stop threshold can be used to prevent the LMedS algorithm iterating too many times in cases where samples have a very similar accuracy. For instance, in cases where proportion of outliers is very small (close to 0%), and samples are very accurate (i.e. 1e-6), the algorithm would iterate for a long time trying to find the best solution when indeed there is no need to do that if a reasonable threshold has already been reached. Because of this behaviour the stop threshold can be set to a value much lower than the one typically used in RANSAC, and yet the algorithm could still produce even smaller thresholds in estimated results.
      See Also:
    • MIN_STOP_THRESHOLD

      public static final double MIN_STOP_THRESHOLD
      Minimum allowed stop threshold value.
      See Also:
    • stopThreshold

      private double stopThreshold
      Threshold to be used to keep the algorithm iterating in case that best estimated threshold using median of residuals is not small enough. Once a solution is found that generates a threshold below this value, the algorithm will stop. The stop threshold can be used to prevent the LMedS algorithm iterating too many times in cases where samples have a very similar accuracy. For instance, in cases where proportion of outliers is very small (close to 0%), and samples are very accurate (i.e. 1e-6), the algorithm would iterate for a long time trying to find the best solution when indeed there is no need to do that if a reasonable threshold has already been reached. Because of this behaviour the stop threshold can be set to a value much lower than the one typically used in RANSAC, and yet the algorithm could still produce even smaller thresholds in estimated results.
    • qualityScores

      private double[] qualityScores
      Quality scores corresponding to each provided sample. The larger the score value the better the quality of the sample.
  • Constructor Details

    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator()
      Constructor.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(List<StandardDeviationBodyKinematics> measurements)
      Constructor.
      Parameters:
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(boolean commonAxisUsed)
      Constructor.
      Parameters:
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] bias)
      Constructor.
      Parameters:
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] initialBias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      initialBias - initial accelerometer bias to be used to find a solution. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope. If true 7 samples are required, otherwise 10.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope. If true 7 samples are required, otherwise 10.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] initialBias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      initialBias - initial accelerometer bias to be used to find a solution. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if quality scores array is smaller than 13 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if quality scores array is smaller than 13 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.ECEFPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if provided quality scores length is smaller than the minimum number of required samples (10 or 13).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope. If true 7 samples are required, otherwise 10.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - list of body kinematics measurements taken at a given position with different unknown orientations and containing the standard deviations of accelerometer and gyroscope measurements.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      listener - listener to be notified of events such as when estimation starts, ends or its progress significantly changes.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, double[] bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, double[] bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias. This must have length 3 and is expressed in meters per squared second (m/s^2).
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias array does not have length 3 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if provided bias matrix is not 3x1 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if quality scores array is smaller than 10 samples.
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
    • PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator

      public PROMedSRobustKnownBiasAndPositionAccelerometerCalibrator(double[] qualityScores, com.irurueta.navigation.frames.NEDPosition position, List<StandardDeviationBodyKinematics> measurements, boolean commonAxisUsed, com.irurueta.algebra.Matrix bias, com.irurueta.algebra.Matrix initialMa, RobustKnownBiasAndPositionAccelerometerCalibratorListener listener)
      Constructor.
      Parameters:
      qualityScores - quality scores corresponding to each provided measurement. The larger the score value the better the quality of the sample.
      position - position where body kinematics measures have been taken.
      measurements - collection of body kinematics measurements with standard deviations taken at the same position with zero velocity and unknown different orientations.
      commonAxisUsed - indicates whether z-axis is assumed to be common for accelerometer and gyroscope.
      bias - known accelerometer bias.
      initialMa - initial scale factors and cross coupling errors matrix.
      listener - listener to handle events raised by this calibrator.
      Throws:
      IllegalArgumentException - if either provided bias matrix is not 3x1 or scaling and coupling error matrix is not 3x3 or if provided quality scores length is smaller than the minimum number of required samples (7 or 10).
  • Method Details

    • getStopThreshold

      public double getStopThreshold()
      Returns threshold to be used to keep the algorithm iterating in case that best estimated threshold using median of residuals is not small enough. Once a solution is found that generates a threshold below this value, the algorithm will stop. The stop threshold can be used to prevent the LMedS algorithm to iterate too many times in cases where samples have a very similar accuracy. For instance, in cases where proportion of outliers is very small (close to 0%), and samples are very accurate (i.e. 1e-6), the algorithm would iterate for a long time trying to find the best solution when indeed there is no need to do that if a reasonable threshold has already been reached. Because of this behaviour the stop threshold can be set to a value much lower than the one typically used in RANSAC, and yet the algorithm could still produce even smaller thresholds in estimated results.
      Returns:
      stop threshold to stop the algorithm prematurely when a certain accuracy has been reached.
    • setStopThreshold

      public void setStopThreshold(double stopThreshold) throws com.irurueta.navigation.LockedException
      Sets threshold to be used to keep the algorithm iterating in case that best estimated threshold using median of residuals is not small enough. Once a solution is found that generates a threshold below this value, the algorithm will stop. The stop threshold can be used to prevent the LMedS algorithm to iterate too many times in cases where samples have a very similar accuracy. For instance, in cases where proportion of outliers is very small (close to 0%), and samples are very accurate (i.e. 1e-6), the algorithm would iterate for a long time trying to find the best solution when indeed there is no need to do that if a reasonable threshold has already been reached. Because of this behaviour the stop threshold can be set to a value much lower than the one typically used in RANSAC, and yet the algorithm could still produce even smaller thresholds in estimated results.
      Parameters:
      stopThreshold - stop threshold to stop the algorithm prematurely when a certain accuracy has been reached.
      Throws:
      IllegalArgumentException - if provided value is zero or negative.
      com.irurueta.navigation.LockedException - if calibrator is currently running.
    • getQualityScores

      public double[] getQualityScores()
      Returns quality scores corresponding to each provided sample. The larger the score value the better the quality of the sample.
      Specified by:
      getQualityScores in interface QualityScoredAccelerometerCalibrator
      Overrides:
      getQualityScores in class RobustKnownBiasAndPositionAccelerometerCalibrator
      Returns:
      quality scores corresponding to each sample.
    • setQualityScores

      public void setQualityScores(double[] qualityScores) throws com.irurueta.navigation.LockedException
      Sets quality scores corresponding to each provided sample. The larger the score value the better the quality of the sample.
      Specified by:
      setQualityScores in interface QualityScoredAccelerometerCalibrator
      Overrides:
      setQualityScores in class RobustKnownBiasAndPositionAccelerometerCalibrator
      Parameters:
      qualityScores - quality scores corresponding to each sample.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than minimum required samples (10 or 13).
      com.irurueta.navigation.LockedException - if calibrator is currently running.
    • isReady

      public boolean isReady()
      Indicates whether solver is ready to find a solution.
      Specified by:
      isReady in interface AccelerometerCalibrator
      Overrides:
      isReady in class RobustKnownBiasAndPositionAccelerometerCalibrator
      Returns:
      true if solver is ready, false otherwise.
    • calibrate

      public void calibrate() throws com.irurueta.navigation.LockedException, com.irurueta.navigation.NotReadyException, CalibrationException
      Estimates accelerometer calibration parameters containing scale factors and cross-coupling errors.
      Throws:
      com.irurueta.navigation.LockedException - if calibrator is currently running.
      com.irurueta.navigation.NotReadyException - if calibrator is not ready.
      CalibrationException - if estimation fails for numerical reasons.
    • getMethod

      public com.irurueta.numerical.robust.RobustEstimatorMethod getMethod()
      Returns method being used for robust estimation.
      Specified by:
      getMethod in class RobustKnownBiasAndPositionAccelerometerCalibrator
      Returns:
      method being used for robust estimation.
    • isQualityScoresRequired

      public boolean isQualityScoresRequired()
      Indicates whether this calibrator requires quality scores for each measurement or not.
      Returns:
      true if quality scores are required, false otherwise.
    • internalSetQualityScores

      private void internalSetQualityScores(double[] qualityScores)
      Sets quality scores corresponding to each provided sample. This method is used internally and does not check whether instance is locked or not.
      Parameters:
      qualityScores - quality scores to be set.
      Throws:
      IllegalArgumentException - if provided quality scores length is smaller than the minimum required number of samples (10 or 13).